Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr.
Fecha de publicación:
Fecha Ahead of Print:
Autores de IDIVAL
Autores ajenos al IDIVAL
- Pitart C
- Marco F
- Rodríguez-Martínez JM
- Pascual A
Abstract
The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6 ')-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted.
Datos de la publicación
- ISSN/ISSNe:
- 0095-1137, 1098-660X
- Tipo:
- Article
- Páginas:
- 3343-3345
- DOI:
- 10.1128/JCM.00563-11
- PubMed:
- 21775549
JOURNAL OF CLINICAL MICROBIOLOGY AMER SOC MICROBIOLOGY
Citas Recibidas en Web of Science: 5
Documentos
- No hay documentos
Filiaciones
Filiaciones no disponibles
Campos de Estudio
Cita
Calvo J, Cano ME, Pitart C, Marco F, Rodríguez JM, Pascual A, Martínez L. Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr. J. Clin. Microbiol. 2011. 49. (9):p. 3343-3345. IF:4,220. (1).