IS THE MONOCLONAL FLUORESCENCE POLARIZATION IMMUNOASSAY FOR CYCLOSPORINE SPECIFIC - COMPARISON WITH SPECIFIC RADIOIMMUNOASSAY

Fecha de publicación:

Autores de IDIVAL

  • Juan Antonio Armijo Simón

    Autor

  • Mª Ángeles De Cos Cossío

    Autor

Autores ajenos al IDIVAL

  • NAVARRO, FA

Abstract

Trough concentrations of cyclosporine (CsA) in whole blood were measured by specific monoclonal H-3-radioimmunoassay (S-RIA) and compared with those obtained by monoclonal fluorescence polarization immunoassay (S-FPIA) in 89 transplant recipients. Differences in the S-RIA:S-FPIA relationship between kidney transplant (KT, n = 59) and heart transplant (HT, n = 24) recipients were investigated. Mean concentrations of CsA were significantly higher by S-FPIA than by S-RIA [191 +/- 127 vs. 166 +/- 124 ng/ml (mean +/- SD), p < 0.001], with a S-FPIA:S-RIA ratio of 1.25 +/- 0.33. Higher ratios were observed either in patients with low CsA concentrations or in those with a high proportion of metabolites. HT recipients had higher S-FPIA:S-RIA ratios than did KT recipients (1.37 vs. 1.22, t = 1.87, p = 0.065). Within-assay coefficients of variation were lower for S-FPIA than for S-RIA (2.4% vs. 12.3%, p < 0.001). Monoclonal FPIA is a precise and fast method, more suitable than RIA for therapeutic CsA monitoring in clinical practice. However, our results indicate a 25% higher mean CsA concentration by S-FPIA than by S-RIA (37% in HT recipients), which should be borne in mind until therapeutic and toxic ranges are established or a more specific FPIA method is developed.

Datos de la publicación

ISSN/ISSNe:
0163-4356, 1536-3694

THERAPEUTIC DRUG MONITORING  LIPPINCOTT WILLIAMS & WILKINS

Tipo:
Article
Páginas:
333-338
PubMed:
1519309

Citas Recibidas en Web of Science: 13

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • CYCLOSPORINE; FLUORESCENCE POLARIZATION IMMUNOASSAY; RADIOIMMUNOASSAY; KIDNEY TRANSPLANT; HEART TRANSPLANT

Cita

Compartir