Long-Term Intracoronary Structural and Vasomotor Assessment of the ABSORB Bioresorbable Vascular Scaffold
Autores de IDIVAL
Autores ajenos al IDIVAL
- Altisent OA
- Goncalves-Ramírez LR
- Fernández L
- Viladés D
- Gutiérrez E
- Mitomo S
- Latib A
- Córdoba-Soriano JG
- Adeliño R
- Amat-Santos I
- Muñoz JF
- Elízaga J
- Bezzera H
- Pereira GTR
- de Prado AP
- Carrillo X
- Fernández-Nofrerias E
- Vilalta V
- Rodríguez-Leor O
- Llibre C
- Fadeuilhe E
- Trujillo A
- Mauri J
- Bayes-Genís A
- Puri R
Unidades
Abstract
We systematically categorized the longer-term (>= 3 years) structural and functional characteristics of the ABSORB bioresorbable vascular scaffold (BVS) using optical coherence tomography imaging and coronary vasomotor reactivity testing and further compared the functional characteristics of BVS stented versus remote coronary segments. A total of 92 patients (mean age 56.4 +/- 9.7 years, 22.8% women) who underwent percutaneous coronary intervention (76% with acute coronary syndrome) using the ABSORB BVS (112 lesions) were included. Optical coherence tomography analysis (38,790 visible struts) comprised insegment quantitative lumen/plaque and semiquantitative plaque composition analysis of the neointimal pattern. Epicardial endothelium-dependent and-independent vasomotion was defined as any vasodilatation at low/intermediate intracoronary dose of acetylcholine (ACh) and nitroglycerine, assessed using quantitative coronary angiography. At a median time of 3.2 years follow-up, 79.8% of BVS segments still demonstrated visible struts with a predominant neointimal fibrotic healing pattern in 84% of BVS segments, with 99.5% of struts demonstrating coverage with apposition. Compared with remote segments, BVS segments demonstrated less endothelium-dependent vasodilatation at low (p = 0.06) and intermediate ACh doses (p = 0.04). Hypertension, longer time interval from index percutaneous coronary intervention, and the degree of in-BVS segment neointimal volume (p <0.03 for all) were each independently associated with abnormal BVS endothelium-dependent vasomotor function. Endothelium-independent function was more likely preserved in non-BVS (remote) segments compared with BVS segments (p = 0.06). In conclusion, at 3+ years post-ABSORB BVS insertion, the rate of complete scaffold resorption was low and residual strut presence was high, with a dominant fibrous healing response contributing toward neointimal hyperplasia and endothelium-dependent and-independent vasomotor dysfunction. (C) 2022 Elsevier Inc. All rights reserved.
Copyright © 2022 Elsevier Inc. All rights reserved.
Datos de la publicación
- ISSN/ISSNe:
- 0002-9149, 1879-1913
- Tipo:
- Article
- Páginas:
- 55-63
- PubMed:
- 35058053
- Enlace a otro recurso:
- www.sciencedirect.com
AMERICAN JOURNAL OF CARDIOLOGY EXCERPTA MEDICA INC-ELSEVIER SCIENCE INC
Citas Recibidas en Web of Science: 5
Documentos
- No hay documentos
Filiaciones
Campos de Estudio
Cita
Altisent OA, Goncalves LR, Fernández L, Viladés D, Gutiérrez E, Mitomo S, Latib A, Córdoba JG, Adeliño R, Amat I, Muñoz JF, Elízaga J, Bezzera H, Pereira GTR, de Prado AP, Carrillo X, Fernández E, Vilalta V, Rodríguez O, Llibre C, Fadeuilhe E, Trujillo A, Mauri J, de la Torre JM, Bayes A, Puri R. Long-Term Intracoronary Structural and Vasomotor Assessment of the ABSORB Bioresorbable Vascular Scaffold. Am. J. Cardiol. 2022. 168. p. 55-63. IF:3,133. (3).
Actividad Investigadora