Validation of Quantitative Flow Ratio-Derived Virtual Angioplasty with Post-Angioplasty Fractional Flow Reserve-The QIMERA-I Study
Fecha de publicación:
Fecha Ahead of Print:
Autores de IDIVAL
Autores ajenos al IDIVAL
- Amat-Santos IJ
- Marengo G
- Sánchez-Luna JP
- Cortés Villar C
- Rivero Crespo F
- Jiménez Díaz VA
- Pérez de Prado A
- Sabaté M
- López-Palop R
- Vegas Valle JM
- Suárez de Lezo J
- Fernandez Cordon C
- Gonzalez JC
- García-Gómez M
- Redondo A
- Carrasco Moraleja M
- San Román JA
Unidades
Abstract
Background: Quantitative flow ratio (QFR) virtual angioplasty with pre-PCI residual QFR showed better results compared with an angiographic approach to assess post-PCI functional results. However, correlation with pre-PCI residual QFR and post-PCI fractional flow reserve (FFR) is lacking. Methods: A multicenter prospective study including consecutive patients with angiographically 50-90% coronary lesions and positive QFR results. All patients were evaluated with QFR, hyperemic and non-hyperemic pressure ratios (NHPR) before and after the index PCI. Pre-PCI residual QFR (virtual angioplasty) was calculated and compared with post-PCI fractional flow reserve (FFR), QFR and NHPR. Results: A total of 84 patients with 92 treated coronary lesions were included, with a mean age of 65.5 +/- 10.9 years and 59% of single vessel lesions being the left anterior descending artery in 69%. The mean vessel diameter was 2.82 +/- 0.41 mm. Procedural success was achieved in all cases, with a mean number of implanted stents of 1.17 +/- 0.46. The baseline QFR value was 0.69 +/- 0.12 and baseline FFR and NHPR were 0.73 +/- 0.08 and 0.82 +/- 0.11, respectively. Mean post-PCI FFR increased to 0.87 +/- 0.05 whereas residual QFR had been estimated as 0.95 +/- 0.05, showing poor correlation with post-PCI FFR (0.163; 95% CI:0.078-0.386) and low diagnostic accuracy (30.9%, 95% CI:20-43%). Conclusions: In this analysis, the results of QFR-based virtual angioplasty did not seem to accurately correlate with post-PCI FFR.
Datos de la publicación
- ISSN/ISSNe:
- 2308-3425, 2308-3425
- Tipo:
- Article
- Páginas:
- -
- DOI:
- 10.3390/jcdd11010014
- PubMed:
- 38248884
Journal of Cardiovascular Development and Disease MDPI
Citas Recibidas en Web of Science: 5
Documentos
- No hay documentos
Filiaciones
Keywords
- coronary physiology; fractional flow reserve; quantitative flow ratio; virtual angioplasty
Campos de Estudio
Cita
Amat IJ, Marengo G, Sánchez JP, Cortés C, Rivero F, Jiménez VA, de la Torre JM, Pérez de Prado A, Sabaté M, López R, Vegas JM, Suárez de Lezo J, Fernandez C, Gonzalez JC, García M, Redondo A, Carrasco M, San Román JA. Validation of Quantitative Flow Ratio-Derived Virtual Angioplasty with Post-Angioplasty Fractional Flow Reserve-The QIMERA-I Study. J. Cardiovasc. Dev. Dis. 2024. 11. (1):14. IF:2,400. (2).
Actividad Investigadora