Characterization of microMOSFET detectors for in vivo dosimetry in high-dose-rate brachytherapy with 192Ir

Fecha de publicación: Fecha Ahead of Print:

Autores de IDIVAL

  • Samuel Ruiz Arrebola

    Autor

  • Rosa Fabregat Borrás

    Autor

  • Maria Jose Fernandez Montes

    Autor

  • Maria Mercedes Perez Macho

    Autor

  • Maria Ferri Molina

    Autor

  • Juan Cardenal Carro

    Autor

  • Maria Teresa Pacheco Baldor

    Autor

  • Javier Tomás Anchuelo Latorre

    Autor

  • Pedro José Prada Gómez

    Autor

Autores ajenos al IDIVAL

  • Rodríguez E
  • García A
  • Tornero-López AM
  • Guirado D

Unidades

Abstract

Purpose The objective of this study was to characterize the Best Medical Canada microMOSFET detectors for their application in in vivo dosimetry for high-dose-rate brachytherapy (HDRBT) with Ir-192. We also developed a mathematical model to correct dependencies under the measurement conditions of these detectors. Methods We analyzed the linearity, reproducibility, and interdetector variability and studied the microMOSFET response dependence on temperature, source-detector distance, and angular orientation of the receptor with respect to the source. The correction model was applied to 19 measurements corresponding to five simulated treatments in a custom phantom specifically designed for this purpose. Results The detectors (high bias applied in all measurements) showed excellent linearity up to 160 Gy. The response dependence on source-detector distance varied by (8.65 +/- 0.06)% (k = 1) for distances between 1 and 7 cm, and the variation with temperature was (2.24 +/- 0.05)% (k = 1) between 294 and 310 K. The response difference due to angular dependence can reach (10.3 +/- 1.3)% (k = 1). For the set of measurements analyzed, regarding angular dependences, the mean difference between administered and measured doses was -4.17% (standard deviation of 3.4%); after application of the proposed correction model, the mean difference was -0.1% (standard deviation of 2.2%). For the treatments analyzed, the average difference between calculations and measures was 4.7% when only the calibration coefficient was used, but it is reduced to 0.9% when the correction model is applied. Conclusion Important response dependencies of microMOSFET detectors used for in vivo dosimetry in HDRBT treatments, especially the angular dependence, can be adequately characterized by a correction model that increases the accuracy of this system in clinical applications.

© 2020 American Association of Physicists in Medicine.

Datos de la publicación

ISSN/ISSNe:
0094-2405, 2473-4209

MEDICAL PHYSICS  WILEY

Tipo:
Article
Páginas:
2242-2253
PubMed:
32031263

Citas Recibidas en Web of Science: 7

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • MOSFET; brachytherapy; correction model; in vivo dosimetry