Successful Retrieval of an Irretrievable Jugular Tesio Catheter Using a Fogarty Arterial Embolectomy Catheter

Fecha de publicación: Fecha Ahead of Print:

Autores de IDIVAL

  • Mª Elena Arnáiz García

    Autor

  • Ana Mª Arnaiz García

    Autor

  • Javier Arnaíz García

    Autor

  • Víctor Expósito García

    Autor

  • Juan Francisco Nistal Herrera

    Autor

  • Felipe José Rodríguez Entem

    Autor

  • Juan José Olalla Antolín

    Autor

Autores ajenos al IDIVAL

  • Gutierrez-Diez, F
  • Lopez-Rodriguez, J
  • Gonzalez-Santos, JM

Unidades

Abstract

Long life expectancy and wide development of therapies have increased the number of patients under artificial treatment for lost kidney function or dialysis. Different options for vascular access are suitable for receiving this therapy. The use of tunneled catheters has consequently increased complications related to its use. A difficult retrieval of catheters caused by a hard fibrin sheath along its trajectory is a common drawback. Herein, we report a woman with suspicion of hemodialysis catheter infection and an irretrievable Tesio catheter. A novel technique using a Fogarty arterial catheter allowed a successful retrieval and avoided an aggressive management.

Datos de la publicación

ISSN/ISSNe:
1538-5744, 1938-9116

VASCULAR AND ENDOVASCULAR SURGERY  SAGE PUBLICATIONS INC

Tipo:
Article
Páginas:
349-351
PubMed:
24407507

Citas Recibidas en Web of Science: 4

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • retrieval; venous; Tesio; catheter; stuck; embolectomy; Fogarty

Compartir